Aztec 16 Instrukcje Operacyjne

Przeglądaj online lub pobierz Instrukcje Operacyjne dla Klucze udarowe Aztec 16. Aztec 16 Operating instructions Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 20
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 0
Operating instructions OI/ADS430–EN Rev. A
Aztec ADS430
Optical dissolved oxygen probe
Measurement made easy
Introduction
The ADS430 (RDO
®
PRO-X) probe is a rugged, reliable
instrument designed to deliver accurate dissolved oxygen (DO)
data across a wide measurement range using the latest optical
technology for DO measurement.
The probe is designed for use with the ABB AWT440
multi-input transmitter that features 'hot-plugging' capability.
'Hot-plugging' enables new or replacement sensors to be
connected without the need to power down the transmitter.
When used in conjunction with the optional ADS430 EZCLEAN
compressed air supply unit (connected to an ABB adaptor
nozzle or flowcell), a schedule can be configured to provide
automatic in situ cleaning of the probe.
For more information
Publications for the associated Aztec AWT440 transmitter are
available for free download from www.abb.com/measurement
(see links and reference numbers below) or by scanning this
code:
search for or click on:
Aztec AWT440 multi-input
transmitter
Commissioning instructions
CI/AWT440-EN
Aztec AWT440 multi-input
transmitter
Operating instructions
OI/AWT440-EN
Aztec ADS430 EZCLEAN
compressed air supply unit
Operating instructions
OI/ADS430/EZCLN-EN
Przeglądanie stron 0
1 2 3 4 5 6 ... 19 20

Podsumowanie treści

Strona 1 - Aztec ADS430

Operating instructions OI/ADS430–EN Rev. A Aztec ADS430Optical dissolved oxygen probeMeasurement made easyIntroductionThe ADS430 (RDO® PRO-X) probe is

Strona 2 - Contents

10 OI/ADS430–EN Rev. A | Aztec ADS430 | Optical dissolved oxygen probeSalinity Correction Required when monitoring water containing high quantities o

Strona 3 - 1 Health & Safety

Aztec ADS430 | Optical dissolved oxygen probe | OI/ADS430–EN Rev. A 116 CalibrationThis section describes how to calibrate the probe and involves me

Strona 4 - 2 System overview

12 OI/ADS430–EN Rev. A | Aztec ADS430 | Optical dissolved oxygen probe6.2 1-Point calibration (water-saturated air)Referring to Fig. 6.1:1. Remove st

Strona 5 - 3 Installation

Aztec ADS430 | Optical dissolved oxygen probe | OI/ADS430–EN Rev. A 1311. If 1-Point Cal is displayed, go to step 14.If 1-Point Cal is not displayed

Strona 6

14 OI/ADS430–EN Rev. A | Aztec ADS430 | Optical dissolved oxygen probe6.3 2-Point calibration (100 % and 0 % saturation)1. Perform a 1-point calibrat

Strona 7 - Exit Select

Aztec ADS430 | Optical dissolved oxygen probe | OI/ADS430–EN Rev. A 1511. Use the / keys to select the probe to be calibrated and press the ke

Strona 8 - 5Sensor setup

16 OI/ADS430–EN Rev. A | Aztec ADS430 | Optical dissolved oxygen probe21. Press the key (below the Continue prompt).The Calibrate page is displayed

Strona 9 - Sensor Setup

Aztec ADS430 | Optical dissolved oxygen probe | OI/ADS430–EN Rev. A 177 Maintenance7.1 Cleaning the sensor cap1. Isolate the transmitter and remove

Strona 10

18 OI/ADS430–EN Rev. A | Aztec ADS430 | Optical dissolved oxygen probe8 SpecificationSensor typeOptical (luminescent) dissolved oxygen sensorProbeIP

Strona 11 - 6 Calibration

Aztec ADS430 | Optical dissolved oxygen probe | OI/ADS430–EN Rev. A 199 Spares and accessories9.1 Spares9.2 Accessories*See Section 3.5, page 6.Ackn

Strona 12 - 1-Point Cal

2 OI/ADS430–EN Rev. A | Aztec ADS430 | Optical dissolved oxygen probeContents1 Health & Safety ...

Strona 13 - Slope 1.054

Contact usOI/ADS430–EN Rev. A 04.2015ABB LimitedProcess AutomationHoward Road, St. NeotsCambridgeshire, PE19 8EUUKTel: +44 (0)870 600 6122Fax: +44 (0

Strona 14 - Calibrate

Aztec ADS430 | Optical dissolved oxygen probe | OI/ADS430–EN Rev. A 31 Health & Safety1.1 Document symbolsSymbols that appear in this document a

Strona 15 - 2-Point Cal

4 OI/ADS430–EN Rev. A | Aztec ADS430 | Optical dissolved oxygen probe2 System overviewADS430 probe components are shown in Fig. 2.1:Fig. 2.1 ADS430 p

Strona 16 - Slope 0.954

Aztec ADS430 | Optical dissolved oxygen probe | OI/ADS430–EN Rev. A 53 Installation3.1 Siting3.2 Probe dimensions3.3 Storing the sensor cap3.4 Fitti

Strona 17 - 7 Maintenance

6 OI/ADS430–EN Rev. A | Aztec ADS430 | Optical dissolved oxygen probe3.5 Mounting / cleaning optionsProbe mounting / cleaning options are shown in Ta

Strona 18 - 8 Specification

Aztec ADS430 | Optical dissolved oxygen probe | OI/ADS430–EN Rev. A 74 Sensor setup – first-time installationTo perform a first-time installation (E

Strona 19 - Acknowledgements

8 OI/ADS430–EN Rev. A | Aztec ADS430 | Optical dissolved oxygen probe5Sensor setup1. Connect the ADS430 probe to the transmitter’s EZLink connector –

Strona 20 - 3KXA494430R4201

Aztec ADS430 | Optical dissolved oxygen probe | OI/ADS430–EN Rev. A 95.1 Sensor SetupUsed to set the tag, measurement units, operational range and c

Komentarze do niniejszej Instrukcji

Brak uwag